/ Catalog / Sweat Analysis Systems, Osmometers, Blood Gas, Hematology, Clinical Chemistry, Urinanalysis, Immunodiagnostics / Clinical Chemistry / ADVIA® 1650 Chemistry System
ADVIA® 1650 Chemistry System

|
Power - Reliability - Expertise |
|
Microvolume technologyThe ADVIA® 1650 has minimal requirements for sample volume (as low as 2 µL/test), reagent consumption (as low as 80 µL/test) and water consumption (25 L/hour) - thus maximizing management efficiency.
Outstanding productivityA rapid, 3-second cycle and system throughput of up to 1650 tests/hour provide powerful high-speed performance, ensure workflow efficiency through peak workloads. Optional universal rack handler facilitates continuous sample feeding for further productivity gains.
Unparalleled walkaway capabilityOn-board capacity of 200 samples and 32,000 tests - plus our exclusive Sample Saver Technology for automatic repeat, dilution and reflex testing without operator intervention - optimizes walkaway time, frees you to work on additional tasks.
Expertise in automationFor 50 years we have been recognized for our innovation - just as clinical chemistry has been recognized as the driver of automation. The tradition continues. The ADVIA 1650 can be integrated with other testing disciplines via the ADVIA WorkCell™ or ADVIA LabCell® automated sample transportation systems.
Technical specifications |
| Overview |
|
| Type of System |
Random, continuous access, batch, discrete processing |
| Throughput Rate |
1650 tests/hour, 1200 tests/hour colorimetric, 450 tests/hour ISE |
| Methods Capacity Onboard |
49 including 3 ISE | | |
|
|
| Sample Handling |
|
| Sample Tubes |
2mL, 5 mL, 7 mL, 10 mL sample cups, microcontainer tubes |
| Sample Tray |
84 samples, positive sample identification |
| Sample Rack Handler (optional) |
Universal 5-position rack, 425 total Onboard sample capacity; continous feed |
| Sample Integrity Check |
Qualitative check for Hemolysis, Lipemia and Icterus |
| STAT Positions |
84, not dedicated |
| Bar Codes |
2 of 5, code 39, code 128, Codabar (NW7) |
| Auto-repeat |
Automatic repeat testing from the retained pre-diluted sample or original sample |
| Auto-dilution |
Automatic dilution from retained pre-diluted sample |
| Auto-reflex Testing |
Automatic ability to perform 3 additional tests based on results of first test |
| Primary Probe |
Liquid Level Sensing, Crash Protection, Clot/Clog Detection and Short Sample Detection |
| Secondary Probe |
Liquid Level Sensing | | |
|
|
| Microvolume Technology |
|
| Automatic Sample Pre-dilution |
Samples diluted 1:5 (30 µL sample + 120 µL saline generates up to 15 test results) |
| pre-dilution tray |
120 dilution cups |
| Original Sample Volume |
2 to 30 µL; average of 2-3 µL per test |
| Average Reagent Volume |
80-120 µL per test |
| Storage Capacity Onboard |
20,000 tests average; 32,200 tests maximum | | |
|
|
| Reaction Area |
|
| Reaction Tray |
221 reusable plastic cuvettes |
| Cuvette Optical Path Length |
10 mm |
| Reaction Bath |
Inert fluorocarbon oil circulation system, 37°C |
| Photometer |
14 fixed wavelengths (340, 410, 451, 478, 505, 545, 571, 596, 658, 694, 751, 805, 845 and 884) |
| Light Source |
12V, 50W halogen lamp, cooled by forced water circulation |
| Assay Methods |
Endpoint, rate reaction, 2-point rate, multi-point homogeneous immunoassay |
| Reaction Times |
3, 4, 5, 10, 15 and 20 minutes |
| Automatic Correction |
Serum blank, cell blank, measurement point change, sample volume change in reassay |
| Point Forwarding |
Automatically extends linearity over assay-range samples | | |
|
|
| Reagent Handling |
|
| Reagent Tray |
2 trays, 50 positions each, refrigerated |
| Reagent Capacity Onboard |
46 colorimetric methods |
| Dispensing System |
2 probes with Liquid Level Sensing |
| Reagent Wedges |
20, 40, 70 mL barcode labeled wedges |
| Onboard Stability |
Up to 60 days |
| Reagent Inventory Management |
Tracks tests remaining, lot number, Onboard stability and expiration date |
| Reagent Dilutions |
Capability to dilute concentrated reagents Onboard | | |
|
|
| Open System Capibility |
|
| Channels |
100 channels; includes user-defined applications |
| Bayer 3rd Party Applications |
Varies by country, can be configured on system | | |
|
|
| ISE |
|
| ISE |
Indirect simultaneous measurement of Na+, K+, Cl- |
| Sample Volume |
28 µL for all three tests |
| Priming |
Automatic priming cycle |
| Electrode Shelf Life |
30,000 Samples |
| Throughput Rate |
450 tests/hour; 150 tubes/hour | | |
|
|
| Calibration/QC |
|
| Calibration Interval |
Up to 60 days |
| Auto-Calibration/QC |
User-defined time interval or with new reagent container |
| View Calibration/QC Data |
Graphical display of calibration curve and QC; real-time QC |
| Calibration/Control Tray |
61 refrigerated positions for calibrators, controls and diluents | | |
|
|
| Data Management |
|
| Computer |
Windows NT®, flat panel display |
| SystemDocumentation |
Operator manual and method sheets on-line |
| Host Interface |
RS 232C bidirectional |
| Data Storage |
70,000 patient samples |
| OnboardMaintenance Logs |
Schedule and monitor routine maintenance activities via software |
| Host Query |
ASCII; system requests work-order or batch of work-orders from host | | |
|
|
| General Specifications |
|
| Power Requirements |
100/110/115/200/220/230 V +/–10%, 26 A, 50/60 Hz, 3 KVA |
| Water Requirements |
Deionized water from a non-pressurized water reservoir. Maximum consumption: 6.8 gallons per hour (25 liters per hour) |
| Drain Requirements |
Minimum of 6.8 gallons per hour (25 liters per hour) |
| Dimensions |
ADVIA 1650:44.45 (h) x 58.27 (w) x 34.13 (d) in., 112.9 (h) x 148.0 (w) x 86.7 (d) cm.ADVIA Rack Handler: 37 (h) x 29 (w) x 41 (d) in.,86.4 (h) x 73.7 (w) x 104.1 (d) cm. |
| Weight |
ADVIA 1650: 1,178 lbs. (534.3 kg)Universal Rack Handler 178 lbs. (80.7 kg) |
| Compliance |
UL, cUL, CE |
| Noise Specifications |
Average with top cover closed 55.91 db |
| Average Heat Output |
4,643 Btu/hr. |
| Operating Temperature Range |
18°C - 30°C/64°F - 86°F |
| Ambient Humidity |
40% - 70% | | |
|
|