CompanyManufactureTrainingContacts
RUS ENG
/ Catalog / Sweat Analysis Systems, Osmometers, Blood Gas, Hematology, Clinical Chemistry, Urinanalysis, Immunodiagnostics / Blood Gas / Rapidlab 1200

Rapidlab 1200

 

Technical specifications

Platform Syringe(1ml)
Min. Vol.
Mode MultiCap
Min. Vol.
Capillaries
Mode
1240 90 ul Full menu 90 ul Full menu
1245 140 ul Full menu 140 ul Full menu
  100 ul pH 35 ul pH
1260 150 ul Full menu 125 ul Full menu
  100 ul pH 95 ul pH
1265 175 ul Full menu 175 ul Full menu
< /TD> 95 ul pH 95 ul pH


Analysis  
Time to Results: 60 seconds or less
QC/AQC  
Analysis Time: 130 seconds automatic QC or ampule
Reagent Cartridge  
Use Life: 30-32 days
Calibration  
1 point
(Approximately 2 minutes):
Every 30 minutes
Full Cal (Approximately 8 minutes): Every 8 hours
Environmental  
Temperature: 15C-32C
Humidity: 5%-85% non-condensing
System Weight  
System Weight 65-68 lbs.
System Dimensions  
Width: 52.0 cm (20.5 in.)
Height: 57.78 cm (22.75 in.)
Depth: 53.34 cm (21 in.)
Weight: 29.5 kg (65 lb)
Power Requirement  
Rating: 150VA
Voltage: 100-240V
Frequency: 50/60Hz
External Interface
USB Printer Port, RS232 Port, RS232 Bar Code Gun Port, RJ10/100 Base T Ethernet Port
Approvals
EN 60601-1-2:2001, Group 1, Class B, CB Test Certificate, UL Certificate, CSA Certified 100V/120V (85V-264V autoranging) 50/60 Hz
Copyright 2001-2003 Holding ANTARES

Tel.: +7 (495) 943-92-54
Tel./Fax: +7 (495) 648 15 71
E-mail: antares@antaresholding.ru 

Development and support 
Prorock Wed-soft