Technical specifications
| Platform | Syringe(1ml) Min. Vol. |
Mode | MultiCap Min. Vol. |
Capillaries Mode |
| 1240 | 90 ul | Full menu | 90 ul | Full menu |
| 1245 | 140 ul | Full menu | 140 ul | Full menu |
| 100 ul | pH | 35 ul | pH | |
| 1260 | 150 ul | Full menu | 125 ul | Full menu |
| 100 ul | pH | 95 ul | pH | |
| 1265 | 175 ul | Full menu | 175 ul | Full menu |
| < /TD> | 95 ul | pH | 95 ul | pH |
| Analysis | |
| Time to Results: | 60 seconds or less |
| QC/AQC | |
| Analysis Time: | 130 seconds automatic QC or ampule |
| Reagent Cartridge | |
| Use Life: | 30-32 days |
| Calibration | |
| 1 point (Approximately 2 minutes): |
Every 30 minutes |
| Full Cal (Approximately 8 minutes): | Every 8 hours |
| Environmental | |
| Temperature: | 15°C-32°C |
| Humidity: | 5%-85% non-condensing |
| System Weight | |
| System Weight | 65-68 lbs. |
| System Dimensions | |
| Width: | 52.0 cm (20.5 in.) |
| Height: | 57.78 cm (22.75 in.) |
| Depth: | 53.34 cm (21 in.) |
| Weight: | 29.5 kg (65 lb) |
| Power Requirement | |
| Rating: | 150VA |
| Voltage: | 100-240V |
| Frequency: | 50/60Hz |
| External Interface | |
| USB Printer Port, RS232 Port, RS232 Bar Code Gun Port, RJ10/100 Base T Ethernet Port | |
| Approvals | |
| EN 60601-1-2:2001, Group 1, Class B, CB Test Certificate, UL Certificate, CSA Certified 100V/120V (85V-264V autoranging) 50/60 Hz | |